Due to its unique combination of submersible probe design and UV fluorescence sensing technology, the FP360 sc delivers the best technology to detect oil in water and is priced below competitive UV fluorescent instruments.
Minimal maintenance
The FP360 sc has no tubes, pumps, or valves that can foul or require constant maintenance interventions. Maintenance is limited to occasional wiping of the sensor's measurement window and a calibration once every two years.
Reduced laboratory testing
While laboratory testing is the ultimate method of measuring oil in water, it is a long and complex process that requires special equipment and trained lab personnel. The FP360 sc provides a cost-effective, continuous online monitoring solution to maintain process control and avoid oil contamination with minimal laboratory testing.
High sensitivity and selectivity
The FP360 sc can detect and measure polycyclic aromatic hydrocarbons (PAHs) from 1.2 ppb to up to 5000 ppb (µg/L). This is approximately equivalent to a concentration of mineral oil between 0.1 to 150 ppm (mg/L). Furthermore, the FP 360 sc method of detection makes it impervious to interferences by turbid water or natural organic and biological matter that impact online light scattering, UV absorbance, and VIS fluorescence instruments.
One to eight sensors
The Hach Digital Controller Family can receive data from up to eight Hach digital sensors, including oil-in-water, suspended solids, turbidity, pH/ORP, dissolved oxygen, conductivity, ammonium, phosphate, SAC, and nitrate in any combination.
There’s no complicated wiring or set up procedures with the family of controllers. Just plug the sensor to any Hach digital controller and it’s ready to use because it’s “plug and play.”
Τι περιέχει το καλάθι
Oil-in-water probe, user manual
Προδιαγραφές
Body Material:Stainless Steel
Αναπαραγωγιµότητα :2.5 % of measured value at constant temperature
Ανιχνευτής :UV photodiode with interference filter; Compensation of daylight and flashlamp intensity fluctuations
Βαθµονόµηση :Factory calibrated with UV fluorescence standard or process calibration with results of a grab sample analysis.